跳到主要内容

LM.2 · Measurement loading, amplifier gain and detector response

GRE 物理知识点 43

本主题的视频课程 打开视频页面
6:08

测量负载、放大器增益与探测器响应

电压表会改变其测量的电路:其读数是一个分压器的输出,而非开路电压。本节计算电表负载效应,设定……

英文讲解 · 可切换中英文字幕

43.1

实验室仪器、测量仪表和探测器

English

In 1928 Hans Geiger and his student Walther Müller, in Kiel, built a gas-filled tube that produced a large electrical pulse for each ionising particle passing through it. The Geiger–Müller counter made counting radiation simple, but it has a weakness: after each pulse the tube is "blind" for about 100 microseconds. At high count rates it misses events, and the true rate must be corrected. Every instrument changes, delays or distorts what it measures in some way, and GRE laboratory questions test whether you can correct for it.

GRE Physics laboratory questions ask you to:

  • Correct a meter reading for loading 负载效应.
  • Find the gains of ideal op-amp circuits.
  • Use the cutoff frequency 截止频率 of an RC filter.
  • Report detector efficiency and energy resolution.
  • Correct a count rate for dead time 死时间.
词汇 训练
English 中文 拼音
loading/ˈləʊdɪŋ/ 负载效应 fù zài xiào yìng
cutoff frequency/ˈkʌtɒf ˈfriːkwənsi/ 截止频率 jié zhǐ pín lǜ
dead time/ded taɪm/ 死时间 sǐ shí jiān
43.2

电表负载效应

English

A voltmeter with resistance $R_m$ across a source with Thévenin resistance $R_{\text{th}}$ forms a voltage divider:

$$V_{\text{read}} = V_{\text{th}}\,\frac{R_m}{R_m + R_{\text{th}}}$$
  • The reading is always low. A meter with $R_m = 10R_{\text{th}}$ reads 90.9% of the true value.
  • A meter's input impedance 输入阻抗 should be much larger than the source resistance. A 10× oscilloscope probe raises it, but also divides the signal by 10.
  • An ammeter loads a circuit the other way: its resistance must be small.
词汇 训练
English 中文 拼音
input impedance/ˈɪnpʊt ɪmˈpiːdəns/ 输入阻抗 shū rù zǔ kàng
43.3

运算放大器与滤波器

English

For an ideal op-amp with negative feedback, the inputs draw no current and sit at the same voltage.

  • Non-inverting gain: $1 + R_2/R_1$. Inverting gain: $-R_2/R_1$.
  • An RC low-pass filter has $f_c = \dfrac{1}{2\pi RC}$.
$$\frac{V_{\text{out}}}{V_{\text{in}}} = \frac{1}{\sqrt{1 + (f/f_c)^2}}, \qquad \phi = -\arctan\frac{f}{f_c}$$
  • At $f_c$ the amplitude ratio is $1/\sqrt2 = 0.707$ (half the power), not $\tfrac12$, and the phase is $-45°$.
Left: meter reading against meter-to-source resistance, 90.9% at 10 R_th. Right: an RC low-pass filter's amplitude ratio, 1/√2 at the cutoff
A finite meter resistance always reads low; a low-pass filter passes 1/√2 at f_c
43.4

探测器

English
  • Efficiency = particles detected ÷ particles reaching the detector.
  • Energy resolution = full width at half maximum ÷ energy. A 13.2 keV width at 662 keV is 2.0%.
  • These are different properties: a detector can count well and still blur energies.
43.5

死时间

English

After each recorded event, a non-paralysable detector cannot record another for a time $\tau$. With true rate $n$, the recorded rate $m$ is:

$$m = \frac{n}{1 + n\tau}, \qquad n = \frac{m}{1 - m\tau}$$
  • The correction matters when $m\tau$ is not small.
43.6

校准和误差

English
  • A linear calibration fixes an offset and a scale. It does not remove noise or nonlinearity.
  • Averaging $N$ readings reduces random error by $\sqrt N$, but leaves a systematic bias unchanged.
43.7

例题:带有倾向性的资料

English
  • A 10 V source with 100 kΩ internal resistance, read by a 1.0 MΩ meter: $10 \times 1.0/1.1 = 9.09$ V.
  • A 12 V source with 200 kΩ, the same meter: $12 \times 1.0/1.2 = 10.0$ V, a 17% error.
  • A non-inverting amplifier with $R_1 = 1.0$ kΩ and $R_2 = 9.0$ kΩ has gain 10.
43.8

例题:筛选器与计数器

English

$R = 1.6$ kΩ and $C = 100$ nF.

  • $f_c = 1/(2\pi \times 1600 \times 1.0 \times 10^{-7}) = 995$ Hz, about 1.0 kHz.
  • At 3.0 kHz: ratio $1/\sqrt{1 + 9} = 0.32$, phase $-\arctan 3 = -72°$.
  • A counter with $\tau = 20$ μs records 8000 per second: $n = 8000/(1 - 0.16) = 9524$ per second.
43.9

常见错误

English
  • Taking the meter reading as the source voltage.
  • Using $\tfrac12$ for the amplitude at the cutoff frequency.
  • Mixing up the inverting and non-inverting gains.
  • Using $m(1 + m\tau)$ instead of $m/(1 - m\tau)$ for dead time.
  • Expecting averaging to remove a systematic error.
43.10

关键术语

English

loading: The change a measuring instrument makes to the circuit it measures.

input impedance: The effective load that a measuring instrument presents to the circuit under test.

cutoff frequency: The frequency at which a filter passes $1/\sqrt2$ of the input amplitude.

dead time: The time after one recorded event during which a detector cannot record another.

43.11

自我检测

English
  • A meter with $R_m = 4R_{\text{th}}$ reads a 5.0 V source. What does it show?
  • What is the phase shift of an RC low-pass filter at its cutoff frequency?
  • A detector with $\tau = 10$ μs has a true rate of $10^4$ per second. What rate does it record?

Answers: 4.0 V. $-45°$. $10^4/1.1 = 9.1 \times 10^3$ per second.

这有帮助吗?

该知识点的互动课程

逐步完成,配合即时检查练习。

更多 GRE 物理 知识点

登录或创建账户