The Beer-Lambert Law
Diperkenalkan| English | Makna |
|---|---|
| spectrophotometer | spectrophotometer |
| calibration line | calibration line |
| absorbance/əbˈsɔːbəns/ | Absorbansi |
Measuring concentration with light
- A deep blue copper sulfate solution is more concentrated than a pale one. A spectrophotometer 分光光度计 turns "how blue" into a number, so you can find a concentration without any titration.
- AP labs use this often, and free-response questions ask about calibration lines, dilution and experimental errors.
The Beer–Lambert law
- Absorbance 吸光度 $A=\varepsilon bc$: $\varepsilon$ is the molar absorptivity (how strongly the substance absorbs at that wavelength), $b$ the path length (usually 1 cm), $c$ the molar concentration.
- With $\varepsilon$ and $b$ fixed, $A$ is directly proportional to $c$: double the concentration, double the absorbance.

The Beer-Lambert law
Absorbance is proportional to concentration - the basis of a calibration curve.
A solution with c = 0.10 M has A = 0.36. What is A for 0.25 M of the same substance, in the same cuvette and at the same wavelength?
Kesulitan 2/5 · 140 koin + bonus kecil pada penyelesaian pertama hari ini
A is proportional to c, so 0.36 × 2.5 = 0.90.
Choosing the wavelength
- A solution absorbs the colour complementary to the one you see: blue copper(II) solutions absorb orange-red light, near 620–650 nm.
- Set the instrument to $\lambda_{\max}$, the wavelength of strongest absorbance. This gives the steepest calibration line and the smallest error.
Which wavelength should be chosen to measure a solution's concentration?
Kesulitan 1/5 · 100 koin + bonus kecil pada penyelesaian pertama hari ini
λmax gives the largest change in A per change in c, so the smallest error.
Using $A=\varepsilon bc$. A dye has $\varepsilon=1.50\times10^4\ \text{M}^{-1}\text{cm}^{-1}$ at its $\lambda_{\max}$. A sample in a 1.00 cm cuvette reads $A=0.450$.
- $c=\dfrac{A}{\varepsilon b}=\dfrac{0.450}{(1.50\times10^4)(1.00)}=3.00\times10^{-5}$ M.
ε = 2.20 × 10³ M⁻¹ cm⁻¹, b = 1.00 cm, A = 0.550. Find c, giving the answer × 10⁻⁴ M (for 1.0 × 10⁻⁴ type 1.0).
Kesulitan 2/5 · 140 koin + bonus kecil pada penyelesaian pertama hari ini
c = 0.550 ÷ 2200 = 2.50 × 10⁻⁴ M.
Calibration lines
- Usually $\varepsilon$ is not given. Measure $A$ for several standard solutions of known concentration and plot $A$ against $c$: a straight line through the origin with slope $\varepsilon b$.
- Measure the unknown at the same wavelength, then read its concentration from the line (or divide $A$ by the slope).
Calibration and dilution. Standards of $\text{Cu}^{2+}$:
| $c$ / M | 0.020 | 0.040 | 0.060 | 0.080 |
|---|---|---|---|---|
| $A$ | 0.120 | 0.240 | 0.360 | 0.480 |
- Slope $=\dfrac{0.480}{0.080}=6.0\ \text{M}^{-1}$. An unknown reads $A=0.300$, so $c=\dfrac{0.300}{6.0}=0.050$ M.
- The unknown had been made by diluting 10.0 mL of a stock solution to 50.0 mL. The stock is $0.050\times\dfrac{50.0}{10.0}=0.25$ M.
A calibration line has slope 12.5 M⁻¹ (A against c). An unknown reads A = 0.475. Find its concentration in M.
Kesulitan 2/5 · 140 koin + bonus kecil pada penyelesaian pertama hari ini
0.475 ÷ 12.5 = 0.0380 M.
Lab errors, step by step
- Ask: does the error make the measured $A$ too high or too low? Then the calculated $c$ moves the same way, because $c=A/(\varepsilon b)$.
- Water left in the cuvette dilutes the sample → $A$ too low → $c$ too low. Fingerprints or scratches block light → $A$ too high → $c$ too high.
A student rinses the cuvette with water but does not dry it before adding the sample. How does this affect the calculated concentration?
Kesulitan 1/5 · 100 koin + bonus kecil pada penyelesaian pertama hari ini
The water dilutes the sample, so A and the calculated c are both too low.
Fingerprints on the clear sides of the cuvette make the calculated concentration too high.
Kesulitan 1/5 · 100 koin + bonus kecil pada penyelesaian pertama hari ini
True. Less light gets through, so A reads too high and so does c.
- The straight line holds only for fairly dilute solutions. If $A$ is above the highest standard, dilute the sample and measure again.
- Use the same wavelength and the same cuvette for the standards and the unknown.
- Blank the instrument with the solvent first, so the solvent's own absorbance is set to zero.
- $A=\varepsilon bc$; at fixed $\varepsilon$ and $b$, $A\propto c$.
- Measure at $\lambda_{\max}$ (the complementary colour). Calibration line slope $=\varepsilon b$.
- An error that raises $A$ raises the calculated $c$, and vice versa.